Skip to main content
  • Research article
  • Open access
  • Published:

Pharmacokinetics of recombinant human growth hormone administered by cool.click™ 2, a new needle-free device, compared with subcutaneous administration using a conventional syringe and needle

Abstract

Background

Growth hormone (GH) is used to treat growth hormone deficiency (GHD, adult and paediatric), short bowel syndrome in patients on a specialized diet, HIV-associated wasting and, in children, growth failure due to a number of disorders including Turner's syndrome and chronic renal failure, and in children born small for gestational age. Different brands and generic forms of recombinant human growth hormone (r-hGH) are approved for varying indications in different countries. New ways of administering GH are required because the use of a needle and syringe or a device where a patient still has to insert the needle manually into the skin on a daily basis can lead to low adherence and sub-optimal treatment outcomes. The objective of this study was to assess the relative bioavailability of r-hGH (Saizen®, Merck Serono) administered by a new needle-free device, cool.click™ 2, and a standard needle and syringe.

Methods

The study was performed with 38 healthy volunteers who underwent pituitary somatotrope cell down-regulation using somatostatin, according to a randomized, two-period, two-sequence crossover design. Following subcutaneous administration of r-hGH using cool.click™ 2 or needle and syringe, pharmacokinetic parameters were analysed by non-compartmental methods. Bioequivalence was assessed based on log-transformed AUC and Cmax values.

Results

The 90% confidence intervals for test/reference mean ratio of the plasma pharmacokinetic variables Cmax and AUC0-inf were 103.7–118.3 and 97.1–110.0, respectively, which is within the accepted bioequivalence range of 80–125%. r-hGH administered by cool.click™ 2 is, therefore, bioequivalent to administration by needle and syringe with respect to the rate and extent of GH exposure. Treatment using cool.click™ 2 was found to be well tolerated. With cool.click™ 2 the tmax was less (3.0 hours) than for needle and syringe delivery (4.5 hours), p = 0.002 (Friedman test), although this is unlikely to have any clinical implications.

Conclusion

These results demonstrate that cool.click™ 2 delivers subcutaneous r-hGH exposure that is bioequivalent to the conventional mode of injection. The new device has the additional advantage of being needle-free, and should help to increase patient adherence and achieve good therapeutic outcomes from r-hGH treatment.

Peer Review reports

Background

Growth hormone deficiency (GHD) affects both children and adults, and clinical manifestations vary depending on the age of onset [1]. Children present with short stature and low growth rate [2], while adults have altered body composition and metabolism with reduced physical performance [3]. At all ages, quality of life is impaired [4, 5].

For many years, replacement therapy using exogenous human growth hormone (GH) has been used successfully to treat children with GHD [6], and has more recently benefited adult patients with GHD [7]. GH is now produced using recombinant DNA technology [8], and is also used to treat growth failure due to a number of other disorders including Turner's syndrome [911] and chronic renal failure [12], and in children born small for gestational age [13].

Conventional GH therapy for GHD was originally developed as a daily subcutaneous injection using a standard needle and syringe. However, many patients (a large proportion of who are children and adolescents) find that using needles is painful and this provokes fear of the injection procedure, resulting in potential non-adherence and sub-optimal therapy. Efforts have focused on finding alternative means of administering GH to patients. Delivery devices such as pre-filled syringes, manual injector pens, auto-injectors, injectors with hidden needles and needle-free devices have been introduced in an attempt to improve dosing accuracy and flexibility, ease-of-use, convenience, adherence and patient-friendliness [1419]. However, the majority of injections still require manual insertion of the needle into the skin by the patient.

Needle-free devices have been introduced for GH therapy, having already been used for some time to administer insulin to patients with diabetes mellitus [20] although, in the latter case, local reactions may have limited more widespread acceptance. These devices expel the liquid preparation of the hormone through a small disposable nozzle at high pressure so that it is forced through the skin and dispersed in the subcutaneous region. This mode of administration is as effective as a conventional injection [21, 22], but has the added advantage of reduced adverse psychological effects [23].

One of the available needle-free devices is the cool.click™ (cool.click® in the USA), a commercially available device customized and introduced by Serono in 2000 for the purpose of injecting recombinant human growth hormone (r-hGH) with variable dosing and child-friendly ergonomic features [24]. The delivery of r-hGH by cool.click™ is bioequivalent to needle injection of r-hGH [22]. It has been shown that patients using needles and syringes to inject r-hGH had lower adherence (more patients missed over half of their prescribed dose) than those using cool.click™, resulting in significantly reduced growth rates than those who missed fewer doses [25]. Some adults and teenagers have indicated a preference for this needle-free injection device, and young children overall favoured it [22, 23], reporting that cool.click™ delivery creates less discomfort than traditional needles.

In response to feedback regarding a wish for simplification of dose selection and improvement of the ergonomics of the present device, the next-generation cool.click™ 2 needle-free injection device for administration of r-hGH has now been developed by Merck Serono (an affiliate of Merck KGaA, Darmstadt, Germany). The new device is similar to the current version of cool.click™, with the additional benefit that it allows dosing in milligrams. The original cool.click™ device allowed dosing only by volume, which meant clinicians had to convert from mass (milligrams r-hGH prescribed) to volume (millilitres of solution to be injected), a procedure that could be further complicated by the fact that different volumes of solvent could be used during reconstitution of the Saizen® powder for injection. In addition, the reading of the cool.click™ linear analogue dosing scale could be difficult – a vertical scale had to be aligned with a horizontal scale to set the required injection volume. In cool.click™ 2, this analogue scale has been replaced by a digital LCD dose readout. Lastly, compared with the original device, cool.click™ 2 is quieter in operation and has a modified design for ease of use and to facilitate handling by children (with smaller hands).

The main objective of this study (Study No. 25821) was to demonstrate that r-hGH administration using the cool.click™ 2 needle-free delivery device was bioequivalent to injection with a standard syringe and needle, the reference standard mode of injection.

Methods

Subjects

Healthy male volunteers with pituitary somatotrope cell down-regulation were screened for eligibility, for recruitment into the study. To be eligible for inclusion, subjects were required to fulfil the following criteria: age 21–50 years; have a body weight greater than 60 kg and a body mass index (BMI) in the range of 22–30 kg/m2; have vital signs in the normal range; and must have agreed to use barrier contraception during the study and for 3 months following completion of the post-study visit. A subject was not entered into the study if he had evidence of any surgical or medical condition that might have interfered with the pharmacokinetics of the investigational medicinal product or if he had received any investigational drug in the 12 weeks prior to dosing.

Study design

The study was designed as a phase I, randomized, open-label, two-period, two-sequence crossover study. Treatment started within 21 days of screening. Each study period lasted 3 days, with a washout period of at least 7 days between drug administrations. The subjects were randomly assigned to one of two treatment sequences. Subjects were allocated a randomization number in sequential, chronological order immediately prior to first dose administration, in accordance with the randomization list supplied by the sponsor (Serono).

The first treatment sequence received a 0.5 mL (2.92 mg) subcutaneous dose of r-hGH (Saizen®, Merck Serono) administered by standard needle and syringe (period 1) followed by administration of the same dose of rhGH using the cool.click™ 2 needle-free injection device (period 2). The second treatment sequence received 0.5 mL (2.92 mg) r-hGH administered by the cool.click™ 2 device (period 1) followed by administration of the same dose of r-hGH using a standard needle and syringe (period 2).

The protocol was approved by the local research ethics committee and conducted in accordance with the Declaration of Helsinki and good clinical practice. Subjects gave written informed consent to participate in the study.

Experimental procedures

The subjects remained in the clinical unit from 16 hours before dosing until 30 hours post-dose. To down-regulate endogenous GH sufficiently to enable accurate assessment of serum GH concentration-time profiles, somatostatin (3 mg) was given intravenously by continuous infusion for 25 hours (corresponding to a rate of approximately 1.75 μg/kg body weight/hour), commencing 1 hour prior to dosing with r-hGH to allow pituitary somatotrope cell down-regulation to be established.

Subcutaneous injections of GH were administered alternately to the left or right lower external abdominal wall with the subject in a relaxed sitting position. A different location on the external abdominal wall was used for the cool.click™ 2 needle-free injection device. The abdominal wall below the umbilicus was divided into two areas; one injection was to be administered in each area. The second injection had to be administered at least 10 cm from the first one. Each injection site was clearly circled with a permanent marker prior to dosing.

The 0.5 mL (2.92 mg) dose of r-hGH administered yielded serum hGH concentrations that remained above the limit of quantification of the hGH assay (Euro/DPC Ltd., UK; lower limit of quantification = 3.1 mIU/L) for a sufficient period to enable accurate assessment of the serum hGH concentration-time profile. Blood samples for determination of PK serum hGH concentrations were taken immediately prior to dosing and at 1, 2, 3, 4, 4.5, 5, 5.5, 6, 7, 8, 10, 12, 18 and 24 hours post-dosing in both treatment periods.

Safety data

All clinical laboratory data outside the normal range were identified. Subject parameters including demographics and baseline characteristics, vital signs and clinical laboratory blood parameters were tabulated and assessed by descriptive statistical analyses.

The post-study examination was performed at the end of the study period 14 ± 3 days after the last dosing. Descriptive summaries were recorded for selected parameters (including demographics and baseline characteristics) using summary statistics [n, mean, standard deviation (SD), median, minimum, maximum] and frequency distributions (n, %).

Close monitoring of adverse events (AEs) was conducted throughout the study, and AEs were recorded but were not statistically evaluated. Local tolerability was assessed by inspection of the injection site at pre-scheduled time points at each period for any local reaction (redness, swelling, induration or bruising) and the severity of the reactions was evaluated.

Data management and analysis methods

Serum concentrations of GH were analysed for each subject by non-compartmental methods using WinNonLin® Professional 4.1 (Pharsight, USA).

The following pharmacokinetic parameters were computed: area under the serum concentration-time curve from time zero to the last quantifiable concentration (AUC0-last); area under the serum concentration-time curve extrapolated to infinity (AUC0-inf); peak serum concentration (Cmax); time of peak serum concentration (tmax); and elimination half-life (t1/2).

The areas under the GH concentration-time curves were calculated according to the log-linear trapezoidal rule [26].

Bioequivalence was assessed according to EU Guideline CPMP/EWP/QWP/1401/98 and the FDA Code of Federal Regulations. Following logarithmic transformation, an analysis of variance (ANOVA, SAS®) was performed on GH metrics (Cmax, AUC0-last, AUC0-inf and tmax) of the full analysis population. There were no imputations for missing data. The ANOVA model consisted of the logarithmically transformed Cmax parameter as the response variable with factors for sequence, subject nested in sequence, period and mode of administration (treatment). Using an average bioequivalence approach, a 90% confidence interval (CI) for the true ratio test (needle-free device) to reference (needle injection) of the means of the two treatments was produced from this model and compared with the equivalence acceptance limits 80–125%.

Based on data from previous Serono r-hGH studies, when the sample size in each sequence group is 15 (and the total sample size is 30), a crossover design has a 90% power to demonstrate equivalence within the acceptance limits of 80–125%, assuming that the expected ratio of means was 1.000, the crossover ANOVA, MSE (ln scale) was 0.250 [the SD differences, σd (ln scale) were 0.354], that data were analysed in the natural log scale using t-tests for differences in means, and that each t-test was made at the 5% level. Taking into account a potential drop-out rate of approximately 20%, it was estimated that approximately 38 subjects were required to complete this study.

The pharmacokinetic analysis population consisted of all 38 subjects (100%) who were randomized into this study and who had evaluable pharmacokinetic data for both periods.

The analysis of the parameter tmax was conducted using the non-parametric Friedman test using untransformed tmax data.

Results

Thirty-eight healthy male volunteers completed the study. Demographic and baseline characteristics for each subject (Table 1) were in compliance with specific inclusion and exclusion criteria. There were no major protocol deviations, no subjects dropped out and no subjects were withdrawn.

Table 1 Summary of baseline subject demographic data

The ANOVA model assumptions were met satisfactorily and there was no significant sequence effect (p = 0.980). The mean ± SD serum concentration vs time profiles for GH following administration of 2.92 mg of r-hGH by either the needle-free device, cool.click™ 2, or by needle injection were generally similar throughout the 25-hour blood-monitoring period (Figure 1). Geometric mean values for AUC0-inf, AUC0-last and t1/2 were similar between the two administration methods (Table 2). The maximum serum GH concentrations (Cmax) of 18–20 ng/mL were observed 3–4.5 hours (tmax) after drug administration (Table 2, Figure 1).

Table 2 Mean (CV%) pharmacokinetic parameters after subcutaneous administration of 2.92 mg r-hGH/subject by cool.click™ 2 or needle injection
Figure 1
figure 1

Mean hGH (± SD) serum concentration vs time profiles following subcutaneous administration of 2.92 mg rhGH/subject using either the cool.click™ 2 device or standard syringe with needle.

The 90% CIs for the ratio of test (new cool.click™ 2 device) to reference (standard syringe with needle) expressed as a percentage for AUC0-last (98.1, 112.0), AUC0-inf (97.1, 110.0) and Cmax (103.7, 118.3) were all within the specified acceptance range (80–125%) for average bioequivalence (Table 2).

The median tmax following dosing with the needle and syringe was 4.5 hours, compared with 3 hours following dosing using the cool.click™ 2 device (Table 2), and Friedman testing showed that this difference was statistically significant (p = 0.002).

No serious or life-threatening AEs were observed, and no subject was withdrawn due to AEs. The majority of AEs were of mild intensity and short duration, most frequently headache (needle and syringe, n = 11 [20%]; cool.click™ 2, n = 9 [13.2%]), and nausea of mild severity (needle and syringe, n = 10 [18.2%] and cool.click™ 2, n = 12 [17.6%]). Although there was a higher incidence of local redness in subjects after use of the cool.click™ 2 device compared with the use of needle and syringe (14 vs 3 affected subjects, respectively), this was generally mild and was not associated with any significant difference in pain, bruising, swelling, induration or itching. Redness after the use of cool.click™ 2 was most prevalent between the 5 minute and 4 hour assessments, but was experienced at each of the scheduled assessments. The incidence of redness was highest at the 5 minute and 4 hour assessments (14 subjects (36.8%) and 9 subjects (23.7%), respectively). Five subjects experienced pain after dosing with needle and syringe and 3 experienced pain after dosing with cool.click™ 2.

In this study, statistical assessment of the extent and rate of absorption of r-hGH shows that administration by a new needle-free device, cool.click™ 2, is bioequivalent to standard needle injection and demonstrates similar good tolerability.

Discussion

The results of this study demonstrate that the concentration-time profile of hGH following subcutaneous delivery of r-hGH using the cool.click™ 2 needle-free injection device is comparable to the concentration-time profile of the same dose of r-hGH administered subcutaneously by a needle and syringe. Regulatory guidance stipulates that the 90% CIs for the ratios (test to reference) of the areas under the serum concentration vs time curves (AUC ratio) and the maximum plasma drug concentrations (Cmax ratio) must fall between 80% and 125% [27], and this study shows that both the rate and extent of exposure of r-hGH meet the accepted criteria for bioequivalence. These criteria have been used for many years, and are the same for all drugs and routes of administration.

For the purposes of establishing bioequivalence, cool.click™ 2 was compared with needle and syringe delivery, which is considered to be the standard reference for bioequivalence assessment of GH therapies. Its predecessor, cool.click™, was previously proven to be bioequivalent to this reference using similar methods [22], and this approach has also been used for another jet injection device [28].

The earlier time to maximum plasma concentration (tmax) for the cool.click™ 2 device compared with needle and syringe delivery is unlikely to have any clinical implications on the chronic dosing regimens used for GH therapy. This is because the r-hGH therapeutic effect is not directly related to tmax as is often the case with a single-dose therapy, for example a hypnotic used to induce sleep. A possible explanation for the difference in tmax might be the nature of the jet injection, which tends to administer the drug deeper than by needle and with a wider spread and, therefore, the hormone is dispersed faster and absorbed into the blood faster. A similar observation has also been recorded with another jet injection device [28].

Although no statistical analysis was performed to compare AEs between the cool.click™ 2 device and needle and syringe methods of administration of r-hGH, both were well tolerated in this study in terms of both AEs and local tolerability and pain assessment. Local redness, generally mild, was the most commonly observed reaction after administration of r-hGH by needle and syringe and by cool.click™ 2. Subjects made no complaints about the device.

Many factors influence patient adherence to GH therapy, and are being addressed by needle-free technology, including the anxiety associated with a 'fear of needles', as well as occupational needle-related injuries. In addition, the functionality and appearance of the delivery device are of key importance for widespread acceptance. The additional aesthetic and ergonomic benefits incorporated into the design of cool.click™ 2 mean that it has the potential to be even more patient-friendly than currently-used devices.

Conclusion

In conclusion, cool.click™ 2 delivers subcutaneous r-hGH exposure that is bioequivalent to exposure following injection by a needle and syringe, but has the additional advantage of being needle-free, making it particularly suitable for children and adults with an aversion to needles. By increasing patient adherence, cool.click™ 2 should help to achieve good therapeutic outcomes for GHD and other disorders that benefit from administration of exogenous GH.

References

  1. Shalet SM, Toogood A, Rahim A, Brennan BM: The diagnosis of growth hormone deficiency in children and adults. Endocr Rev. 1998, 19 (2): 203-223. 10.1210/er.19.2.203.

    Article  CAS  PubMed  Google Scholar 

  2. Preece MA: Diagnosis and treatment of children with growth hormone deficiency. Clin Endocrinol Metab. 1982, 11 (1): 1–24-10.1016/S0300-595X(82)80035-2.

    Article  PubMed  Google Scholar 

  3. Carroll PV, Christ ER, Sonksen PH: Growth hormone replacement in adults with growth hormone deficiency: assessment of current knowledge. Trends Endocrinol Metab. 2000, 11 (6): 231-238. 10.1016/S1043-2760(00)00268-X.

    Article  CAS  PubMed  Google Scholar 

  4. Stabler B: Impact of growth hormone (GH) therapy on quality of life along the lifespan of GH-treated patients. Horm Res. 2001, 56 Suppl 1: 55-58. 10.1159/000048136.

    CAS  PubMed  Google Scholar 

  5. McGauley G: The psychological consequences and quality of life in adults with growth hormone deficiency. Growth Horm IGF Res. 2000, 10 Suppl B: S63–S68-

    PubMed  Google Scholar 

  6. Raben MS: Treatment of a pituitary dwarf with human growth hormone. J Clin Endocrinol Metab. 1958, 18 (8): 901–903-

    Article  PubMed  Google Scholar 

  7. Verhelst J, Abs R, Vandeweghe M, Mockel J, Legros JJ, Copinschi G, Mahler C, Velkeniers B, Vanhaelst L, Van Aelst A, De Rijdt D, Stevenaert A, Beckers A: Two years of replacement therapy in adults with growth hormone deficiency. Clin Endocrinol (Oxf). 1997, 47 (4): 485–494-10.1046/j.1365-2265.1997.3041112.x.

    Article  Google Scholar 

  8. Zeisel HJ, von Petrykowski W, Wais U: Pharmacokinetics and short-term metabolic effects of mammalian cell-derived biosynthetic human growth hormone in man. Horm Res. 1992, 37 (Suppl 2): 5–13-

    Google Scholar 

  9. Plotnick L, Attie KM, Blethen SL, Sy JP: Growth hormone treatment of girls with Turner syndrome: the National Cooperative Growth Study experience. Pediatrics. 1998, 102 (2 Pt 3): 479–481-

    PubMed  Google Scholar 

  10. Pasquino AM, Pucarelli I, Segni M, Tarani L, Calcaterra V, Larizza D: Adult height in sixty girls with Turner syndrome treated with growth hormone matched with an untreated group. J Endocrinol Invest. 2005, 28 (4): 350-356.

    Article  CAS  PubMed  Google Scholar 

  11. Stephure DK: Impact of growth hormone supplementation on adult height in turner syndrome: results of the Canadian randomized controlled trial. J Clin Endocrinol Metab. 2005, 90 (6): 3360-3366. 10.1210/jc.2004-2187.

    Article  PubMed  Google Scholar 

  12. Fine RN, Kohaut EC, Brown D, Perlman AJ: Growth after recombinant human growth hormone treatment in children with chronic renal failure: report of a multicenter randomized double-blind placebo-controlled study. Genentech Cooperative Study Group. J Pediatr. 1994, 124 (3): 374–382-

    Article  PubMed  Google Scholar 

  13. Dahlgren J, Wikland KA: Final height in short children born small for gestational age treated with growth hormone. Pediatr Res. 2005, 57 (2): 216–222-10.1203/01.PDR.0000148716.71231.81.

    Article  PubMed  Google Scholar 

  14. Fidotti E: A history of growth hormone injection devices. J Pediatr Endocrinol Metab. 2001, 14 (5): 497-501.

    Article  CAS  PubMed  Google Scholar 

  15. Sjoblom K, Albertsson-Wikland K, Bengtsson BA, Johannsson G, Thoren M, Degerblad M, Savage MO: Patient evaluation of a new injection pen for growth hormone treatment in children and adults. Acta Paediatr. 1995, 411: 63–65-

    Google Scholar 

  16. Verhagen A, Ebels JT, Dogterom AA, Jonkman JH: Pharmacokinetics and pharmacodynamics of a single dose of recombinant human growth hormone after subcutaneous administration by jet-injection: comparison with conventional needle-injection. Eur J Clin Pharmacol. 1995, 49 (1-2): 69–72-10.1007/BF00192361.

    Article  PubMed  Google Scholar 

  17. Bareille P, MacSwiney M, Albanese A, De Vile C, Stanhope R: Growth hormone treatment without a needle using the Preci-Jet 50 transjector. Arch Dis Child. 1997, 76 (1): 65–67-

    Article  PubMed  PubMed Central  Google Scholar 

  18. Silverstein JH, Murray FT, Malasanos T, Myers S, Johnson SB, Frye K, Grossman M: Clinical testing results and high patient satisfaction with a new needle-free device for growth hormone in young children. Endocrine. 2001, 15 (1): 15–17-10.1385/ENDO:15:1:015.

    Article  PubMed  Google Scholar 

  19. Dorr HG, Zabransky S, Keller E, Otten BJ, Partsch CJ, Nyman L, Gillespie BK, Lester NR, Wilson AM, Hyren C, van Kuijck MA, Schuld P, Schoenfeld SL: Are needle-free injections a useful alternative for growth hormone therapy in children? Safety and pharmacokinetics of growth hormone delivered by a new needle-free injection device compared to a fine gauge needle. J Pediatr Endocrinol Metab. 2003, 16 (3): 383–392-

    Article  PubMed  Google Scholar 

  20. Lindmayer I, Menassa K, Lambert J, Moghrabi A, Legendre L, Legault C, Letendre M, Halle JP: Development of new jet injector for insulin therapy. Diabetes Care. 1986, 9 (3): 294-297. 10.2337/diacare.9.3.294.

    Article  CAS  PubMed  Google Scholar 

  21. Houdijk EC, Herdes E, Delemarre-Van de Waal HA: Pharmacokinetics and pharmacodynamics of recombinant human growth hormone by subcutaneous jet- or needle-injection in patients with growth hormone deficiency. Acta Paediatr. 1997, 86 (12): 1301-1307.

    Article  CAS  PubMed  Google Scholar 

  22. Murray FT, Silverstein JH, Johnson SB, Gertner JH, Frye K, Gironda G, Wang Q, Stene M: Bioequivalence and Patient Satisfaction with a Growth Hormone (Saizen®) Needle-Free Device - Results of Clinical and Laboratory Studies. Today's Therapeutic Trends. 2000, 18 (1): 71-86.

    Google Scholar 

  23. Verrips GH, Hirasing RA, Fekkes M, Vogels T, Verloove-Vanhorick SP, Delemarre-Van de Waal HA: Psychological responses to the needle-free Medi-Jector or the multidose Disetronic injection pen in human growth hormone therapy. Acta Paediatr. 1998, 87 (2): 154–158-10.1080/08035259850157589.

    Article  PubMed  Google Scholar 

  24. Cool.click: a needle-free device for growth hormone delivery. Med Lett Drugs Ther. 2001, 43 (1095): 2-3.

  25. Desrosiers P, O'Brien F, Blethen S: Patient outcomes in the GHMonitor: the effect of delivery device on compliance and growth. Pediatr Endocrinol Rev. 2005, 2 Suppl 3: 327-331.

    PubMed  Google Scholar 

  26. Gibaldi M, Perrier D: Pharmacokinetics. 1982, New York , Marcel Dekker, 2nd edition

    Google Scholar 

  27. Birkett DJ: Generics - equal or not?. Aust Prescr. 2003, 26 (4): 85-87.

    Article  Google Scholar 

  28. Agerso H, Moller-Pedersen J, Cappi S, Thomann P, Jesussek B, Senderovitz T: Pharmacokinetics and pharmacodynamics of a new formulation of recombinant human growth hormone administered by ZomaJet 2 Vision, a new needle-free device, compared to subcutaneous administration using a conventional syringe. J Clin Pharmacol. 2002, 42 (11): 1262-1268. 10.1177/009127002762491361.

    Article  CAS  PubMed  Google Scholar 

Pre-publication history

Download references

Acknowledgements

Writing assistance was provided by Joanna Brown. Both the study and preparation of the manuscript were funded by Serono International SA, Geneva, Switzerland.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Michel Christen.

Additional information

Competing interests

CB was and MC is an employee of Merck Serono. AP is and JL-S was an employee of LCG Bioscience, which was engaged by Serono to perform this study.

Authors' contributions

All the authors were involved in the conduct of the study, all critically reviewed and revised the manuscript and all have given final approval of the submitted version.

Authors’ original submitted files for images

Below are the links to the authors’ original submitted files for images.

Authors’ original file for figure 1

Rights and permissions

Open Access This article is published under license to BioMed Central Ltd. This is an Open Access article is distributed under the terms of the Creative Commons Attribution License ( https://creativecommons.org/licenses/by/2.0 ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Reprints and permissions

About this article

Cite this article

Brearley, C., Priestley, A., Leighton-Scott, J. et al. Pharmacokinetics of recombinant human growth hormone administered by cool.click™ 2, a new needle-free device, compared with subcutaneous administration using a conventional syringe and needle. BMC Clin Pharmacol 7, 10 (2007). https://doi.org/10.1186/1472-6904-7-10

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/1472-6904-7-10

Keywords